The Mindray BS-240 is a compact, fully automated, benchtop clinical chemistry analyzer designed to meet the diverse needs of small to medium-sized laboratories. Despite its small footprint, it offers robust functionality and advanced features, making it an optimal solution for laboratories with varying test volumes.
Key Features:
Compact Design: The BS-240 is one of the smallest 200 tests per hour (T/H) throughput automatic chemistry analyzers available, saving valuable laboratory space. It provides 40 positions for samples and 40 positions for reagents, with 40 of these positions being interchangeable to extend capacity as needed.
High Throughput: Capable of processing up to 200 photometric tests per hour, with the option to increase to 400 tests per hour when equipped with an Ion-Selective Electrode (ISE) module.
Efficient Operation: Features a minimum reaction volume of 100 μL, reducing reagent consumption and operational costs.
Advanced Sampling Technology: Incorporates smart-sampling technology that enables automatic hemolysate preparation for Hemoglobin A1c (HbA1c) testing, enhancing productivity and accuracy.
User-Friendly Interface: Equipped with intelligent software and a user-friendly interface, the BS-240 simplifies operation and maintenance, including step-by-step maintenance guides.
Technical Specifications
System function
Automatic, Discrete, Random Access, Bench-top
STAT sample priority
Throughput: Up to 200 tests/hour, up to 400 tests/hour with ISE
Measuring principles: Absorbance photometry, Turbidimetry, Ion
Selective Electrode technology
Methodology: End-point, Fixed-time, Kinetic, optional ISE,
Single/Dual/ reagent chemistries,
monochromatic / bi-chromatic
Original system pack reagent ready to use
Close system and open system is optional
Reagent/Sample Handling
Reagent/Sample tray: 80 positions for reagents and 40 positions
for samples in 24-hour refrigerated
compartment (2~12℃)
Reagent volume: 10~250μl, step by 0.5μl
Sample volume: 2~45μl, step by 0.1μl
Reagent/Sample probe: Liquid level detection, vertical collision
protection and inventory checking, reagent pre-warming
Probe cleaning: Automatic washing for interior and exterior
Carry over < 0.05%
Automatic sample dilution: Pre-dilution and post-dilution
Internal bar code reader (optional)
Used for sample and reagent programming
Be applicable to various bar code systems of Codabar, ITF
(Interleaved Two of Five)、code128、code39、UPC/EAN、
Code93
Capable to communicate with LIS in bi-directional mode
Reaction System:
Reaction rotor: Rotating tray, containing 40 cuvettes
Cuvette: Reusable, optical length 5mm
Reaction volume: 100~360μl
Operating temperature: 37℃
Temperature fluctuation: ± 0.1℃
ISE Module (optional)
Measuring K+, Na+, ClMixing Unit
Independent mixing bar
Cuvette Washing: Washing station with pre-warmed detergent
and water
Optical System
Light Source: Halogen-tungsten lamp
Wavelength: 8 wavelengths, 340nm、405nm、450nm、
510nm、546nm、578nm、630nm、670nm
Absorption range: 0~4.0 Abs (10mm conversion), resolution
0.0001Abs
Stray Light 5.6Abs
Control and Calibration
Calibration modes: Linear (one point, two points and
multi-points), Logit-Log 4P, Logit-Log 5P,
spline, exponential, polynomial, parabola
Control Rules: X-R, L-J, Westgard multi-rule, Cumulative sum
check, twin plot
Operation Unit
Operation system: Windows 8
Interface: RS-232
Working Conditions
Power Supply: 200~240V, 50/60Hz, ≤1000VA or 100~130V,
60Hz, ≤1000VA
Dimension: 690 mm (length) ×580 mm (depth) ×595 mm
(height)
Weight: 79 kg
Water Consumption: ≤ 4 L/H